Auxiliary Module
Flexible Add-on Modules for Control, Measurement, BMS Communication, and More
Neware add-on solutions help laboratories and production teams build a more complete battery testing workflow. By integrating control units, auxiliary voltage and temperature modules, CAN communication boxes, and high-precision internal resistance testers, users can collect more data, improve test safety, and synchronize external signals with battery cycling results.
- Control unit — Centralized test control, data buffering, offline operation
- CA auxiliary unit — Cell voltage and temperature monitoring
- CAN box — BMS communication via CAN / CAN FD / RS485
- Impedence Tester — Internal resistance and impedance measurement
Explore Modules
Control Unit
— Reliable Command Transfer and Offline Data Protection
The control unit acts as the bridge between the computer and Neware battery testers. Test commands are created and sent from the BTS software on the computer, received by the control unit, and then distributed to all testers connected under the same system.
During testing, the control unit continuously manages command transfer, test execution, data collection, and basic calculation of key test results such as time, capacity, and energy. It also supports real-time protection logic, including voltage limits, current limits, step judgment, cutoff judgment, and advanced protection conditions.
With its built-in 1GB data buffer, the control unit can continue operating even if the computer is accidentally shut down or disconnected. It keeps sending commands to the connected testers, receives test data, and temporarily stores the data inside the control unit. Once the computer is reconnected, the stored data will be automatically uploaded and backed up to the computer, helping prevent test interruption and data loss.
Applications
Centralized Tester Control
Offline Test Continuity
Temporary Data Storage (1GB)
Automatic Data Backup
Tester Model



CT-4001-ZWJPLUS-1U
Control Unit Load Capacity
1Hz Sampling Mode
- Max. total channels: 256 channels
- Data recording interval: 1s
10Hz Sampling Mode
- Max. total channels: 80 channels
- Data recording interval: 0.1s
100Hz Sampling Mode (BTS6000)
- Max. total channels: 8 channels
- Data recording interval: 0.01s
*Note: Auxiliary channels are included in the total channel count.
Auxiliary Units
— Synchronized Voltage and Temperature Acquisition
Auxiliary units expand Neware battery testers with additional voltage and temperature measurement channels. They are designed to work together with the main charge and discharge channels, allowing users to monitor extra cell-level signals during battery testing.
During operation, auxiliary voltage and temperature data can be bound to the main channel test records. These signals can be used for test cutoff conditions, safety protection, and data analysis, helping users evaluate battery behavior with more complete measurement information.
Applications
Cell Voltage Monitoring
Multi Temperature Monitoring
Flexible Integration
Safety Protection Conditions
Module Selection




CA-4008n-1U-5VT-TC-DB25
- Measurement: Voltage / Temperature
- Voltage Range: 0V–5V
- Voltage Accuracy: ±0.05% FS
- Temperature Range: -200°C–260°C
- Temperature Accuracy: ±1°C
- Temperature Resolution: 0.1°C
- Sampling Rate: 10Hz
- Communication: RS485
- Protection: Voltage & Temperature Limits
Datasheet




CA-4032n-1U-T-TC-DB25
- Measurement: Temperature
- Temperature Range: -200°C–260°C
- Temperature Accuracy: ±1°C
- Temperature Resolution: 0.1°C
- Channels Per Unit: 32 Channels
- Sampling Rate: 10Hz
- Communication: RS485
- Data Export: Excel / TXT
- Protection: Temperature Upper / Lower Limits
Datasheet
CAN Communication Box
— CAN/CAN FD and RS485 BMS Integration
The CAN Communication Box enables data communication between Neware battery testing systems and external BMS or communication devices. It supports CAN, CAN FD, and RS485 communication, helping users integrate BMS data into the battery testing workflow.
During testing, the communication box can receive data from the connected BMS or external device and transmit it to the Neware system for monitoring, test judgment, and data analysis. It helps connect battery test channels with real-time BMS information, making it suitable for module and pack-level testing applications.
The device supports web-based configuration, including IP settings, communication mode selection, port settings, and CAN/CAN FD baud rate configuration. Users can select different models according to the required number of CAN and RS485 communication ports.
Applications
BMS Communication
CAN/CAN FD Integration
RS485 Data Acquisition
Configurable Port Settings
Tester Model




CAE-6004n / CAE-6008n-BMS-CANFD-RS485
- Communication: CAN / CAN FD / RS485
- Available Models: 4 CAN + 4 RS485 / 8 CAN + 8 RS485
- Configuration: Web-based settings
- Default IP: 192.168.3.123
- Operation Mode: CAN/CAN FD mode or serial communication mode
- CAN FD Settings: Arbitration baud rate / data baud rate
- Port Settings: Target IP and target port configuration
- Typical Use: BMS data communication and system integration
BT-9562
— High-precision Resistance and Impedance Measurement
The BT-9562 is designed for high-precision battery internal resistance and voltage measurement. It supports resistance, voltage, reactance, complex impedance, and impedance phase angle testing, helping users evaluate battery performance, consistency, and quality.
Using the AC four-terminal measurement method at 1kHz, the BT-9562 provides stable and accurate measurement results for cell inspection, incoming material evaluation, production quality control, and laboratory analysis.
The instrument supports multiple external interfaces, including Ethernet, RS-232C, and USB, making it easy to integrate into different testing workflows. It also provides data storage, comparator settings, abnormal measurement detection, self-calibration, and external I/O for automated sorting applications.
Applications
Internal Resistance Testing
Battery Voltage Measurement
Impedance Analysis
Quality Control and Sorting
Tester Model



BT-9562
- Measurement: Resistance / Voltage / Reactance / Complex Impedance / Phase Angle
- Measurement Method: AC 4-terminal method
- Test Frequency: 1kHz ±0.02Hz
- Rated Input Voltage: DC ±60V
- Max. Voltage to Ground: DC ±70V
- Impedance Range: 0–3Ω
- Voltage Range: 0–60V
- Voltage Accuracy: ±0.01%
- Interfaces: 100M Ethernet / RS-232C / USB
- Data Storage: 1,000,000 groups
- Comparator Settings: 1000 groups
- External I/O: 2 inputs / 4 outputs